1D Vertical Electrical Sounding (VES) with LandMapper Procedure

standard big manual VES cable set by LandviserThe technique and procedure described here can be performed with LandMapper ERM-01 or ERM-02 (set in resistivity mode). The electrode spacings provided in this example are identical to Landviser's supplied "big manual VES" cable set made to measure 16 layers of topsoil down to approximately 9 m. The worksheet for pre-set electrode spacings in such cable re-calculating measured resistivities to 1D VES profile can be downloaded as Manual 1D VES workbook (MS Excel format).

Other electrode spacings are possible for custom-made cable arrays to reach deeper profiles. For example, we developed and tested with LandMapper a 60m-long cable, measuring down to ~ 20 m for one custom hydrology project

This manual VES technique is most convenient to use with three people. Follow step-by-step instructions below. If you need further help, do not hesitate to contact Landviser, LLC @ +1-609-412-0555 or info@landviser.com. Register on our site and download 7 related publications and software!simple manual VES technique for 3 people Manual 1D VES workbook

 

  1. On your site establish the center location for the VES profile. It is a center point between M2 and N2 in the diagram above. Here the person with LandMapper will stay and conduct measurements.
  2. Send two persons outward in a straight line from the center to unwind both red wires from the probe up to 14 m mark (last electrode) and insert electrodes in soil => A1 and B1.
  3. Measuring person grounds central black M1N1 electrodes at MN/2=1.0 m (green label, black wires). Connect black wires (MN) to MN- terminals of LandMapper.
  4. Connect red wires (AB) to AB- terminals of LandMapper.
  5. Take ER-reading at K0=1 and AB electrodes at 14 m.
  6. Move A and B electrodes to 12 m marks at both sides - take ER-reading.
  7. Continue  until AB/2=6.0 m.
  8. Move MN electrodes to 0.5 m mark (black wire, green label) and AB to 5.0 m mark.
  9. Continue taken measurements until last AB/2=1.0 m. It is also previous location of M1N1 electrodes for steps 3-7.
  10. Field ER-readings can be stored in LandMapper memory or recorded manually.
  11. The recorded ER-values needs to be multiplied on appropriate K-coefficients. Use tables below or input in Excel file to carry out auto-calculations.

Note: if ER values measured on deep AB/2 (#1-5) are too small, switch to K1=10, or even K2=100 (99.99), don't forget to record which K-taken you used in the table,  then the table on second page in the Excel workbook will re-calculate ER automatically and plot VES profile.

Field data recording table. You can print this table from the website or download  Manual 1D VES workbook Excel fileas there is a table already pre-formatted to fit on Landscape page.

AB

MN

Measurement from deep to shallow

 

 

 

 

distances, m

m

#

AB/2, m

K-set

K-taken

ER-1

ER-2

ER-3

ER-4

[A1B1] = 28

2.0

1

14.0

K0=1

1

10.41

 

 

 

[A2B2] = 24

2.0

2

12.0

K0=1

1

27.47

 

 

 

[A3B3] = 20

2.0

3

10.0

K0=1

1

102.8

 

 

 

[A4B4] = 18

2.0

4

9.0

K0=1

1

15.75

 

 

 

[A5B5] = 16

2.0

5

8.0

K0=1

1

13.49

 

 

 

[A6B6] = 14

2.0

6

7.0

K0=1

1

10.26

 

 

 

[A7B7] = 12

2.0

7

6.0

K0=1

1

21.28

 

 

 

[A8B8] = 10

1.0

8

5.0

K0=1

1

30.07

 

 

 

[A9B9] = 9

1.0

9

4.5

K0=1

1

12.29

 

 

 

[A10B10] = 8

1.0

10

4.0

K0=1

1

13.73

 

 

 

[A11B11] = 7

1.0

11

3.5

K0=1

1

15.37

 

 

 

[A12B12] = 6

1.0

12

3.0

K0=1

1

62.18

 

 

 

[A13B13] = 5

1.0

13

2.5

K0=1

1

16.02

 

 

 

[A14B14] = 4

1.0

14

2.0

K0=1

1

26.04

 

 

 

[A15B15] = 3

1.0

15

1.5

K0=1

1

18.18

 

 

 

[A16B16] = 2

1.0

16

1.0

K0=1

1

18.18

 

 

 

Instead of writing ER values on the paper in the provided table, you can store ER in LandMapper's memory. You still need to keep track of what cell # starts new VES profile and which K values had you used (those K1-K9 are stored indefinitely, but don't change K in the middle of field trip).

Recalculation table. This table automatically recalculates field ER measurements taken in the table above (in Excel file). Here on website it just shows values of K-geometric coefficient (K-multiply) for pre-set distances among electrodes using formulas for Schlumberger and Wenner arrays.







AB MN Measurement from deep to shallow Recalculate ER for VES interpretation
distances, m m AB/2, m K-multiply K-taken ER-1
[A1B1] = 28 2.0 1 14.0 306.3 1 3188.6
[A2B2] = 24 2.0 2 12.0 224.6 1 6170.4
[A3B3] = 20 2.0 3 10.0 155.5 1 15986
[A4B4] = 18 2.0 4 9.0 125.7 1 1979.2
[A5B5] = 16 2.0 5 8.0 99.0 1 1335
[A6B6] = 14 2.0 6 7.0 75.4 1 773.59
[A7B7] = 12 2.0 7 6.0 55.0 1 1169.9
[A8B8] = 10 1.0 8 5.0 77.8 1 2338.1
[A9B9] = 9 1.0 9 4.5 62.8 1 772.21
[A10B10] = 8 1.0 10 4.0 49.5 1 679.36
[A11B11] = 7 1.0 11 3.5 37.7 1 579.44
[A12B12] = 6 1.0 12 3.0 27.5 1 1709.3
[A13B13] = 5 1.0 13 2.5 18.8 1 301.97
[A14B14] = 4 1.0 14 2.0 11.8 1 306.78
[A15B15] = 3 1.0 15 1.5 6.3 1 114.23
[A16B16] = 2 1.0 16 1.0 2.4 1 42.836

sample VES profile in pseudo-depthsAfter automatic recalculation of VES profile in a table above, Excel file shows 1D profile of electical resistivity in "pseudo-depths". To interprete VES profile to exact depths and layers you may use a number of free 1D resistivity interpetation programs: iVES - SETUP-VES1D.zip, RES1DINV - r1d.zip, IP-Win etc.; most of which can be downloaded from this website.

Good Field Practice: If you wish to relay on stored ER data, make sure to set up data connection to your computer BEFORE you head to the field! You can store some mock data and do download and memory clean up before you go out and start collecting real data! Also clean up ERM-01 memory from all mock data before actual field session, so you don’t get confused which data are those!

Register on our site and download 7 related publications and software!

Locations

San Antonio 29° 25' 26.8392" N, 98° 29' 37.0608" W
Dmitrov 56° 20' 39.0192" N, 37° 31' 2.5716" E

Please to find the details

Please to find the details about custom hydrology project and The article is really very informative and effective. You can check something related to power issues about servo voltage stabilizer

One of the main mistakes that

One of the main mistakes that many people make is using extension cords to permanently provide power to devices that are far away from an outlet. Although this may seem like an entirely safe and normal fix for this problem, it can actually be very hazardous. Whenever there are extension cords running through your house, you run the risk of walking on top of them, which can lead to tripping, pulling things off of surfaces and wearing down the cord itself. As a result, the cord is no longer insulated properly against electrical current. This can lead to overheating, fire, electrical shock and other dangers.

Electrical Contractors Hawaii

Tips on

Pretty awesome post !!! Whatever you wrote here about 1D vertical electric sounding is really effective. Every part of this article seems to me attractive and also vital. Thanks for allocation. :) tips on how to hire good electrician

You have a wonderful comment

You have a wonderful comment pertaining to this post and you have a very useful tip and suggestion on how to hire electricians. I can benefit on this and I can find electrician school in Texas more easier with your suggestions. I think we should collaborate.

This is really great. I'll be

This is really great. I'll be adding this later to my portfolio and set of technological compilations.

Get structural essays at http://www.essaypanda.org/blog/apa-structure now.

Please send me a free under

Please send me a free under ground water electric resistivity survey soft ware

Hi I would like to buy your

Hi I would like to buy your Landmapper for underground water exploration. Can it be used to imager up to 100m down? Cheers HENRI

Dear Henri, LandMapper has

Dear Henri, LandMapper has been tested alongside with conventional geophysical equipment down to 25 m. Theoretically, based on technical specs of the device, there should be no limitations on most soils to 100 m or even deeper. The cables should be about 3 times longer than depth of interest, so doing manual deep VES might be cumbersome and you will need 2 helpers. Right now we are working on creating more convinient semi-manual VES cable set and 2D setup with LandMapper for use by one person. I will post those new developments shortly.

Larisa Golovko (Pozdnyakova), Ph.D. - President of Landviser, LLC

Location

Javascript is required to view this map.